Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
The product contains a hard-coded password, which it uses for its own inbound authentication or for outbound communication to external components.
The product contains hard-coded credentials, such as a password or cryptographic key.
Link | Tags |
---|---|
https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01 | us government resource third party advisory mitigation |