Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
The product stores sensitive information in cleartext within a resource that might be accessible to another control sphere.
The product exposes sensitive information to an actor that is not explicitly authorized to have access to that information.
Link | Tags |
---|---|
https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01 | us government resource third party advisory mitigation |