Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
The product does not perform or incorrectly performs an authorization check when an actor attempts to access a resource or perform an action.
Link | Tags |
---|---|
https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01 | mitigation third party advisory us government resource |