An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0; 3.0; and 3.1. CADD-Solis Medication Safety Software grants an authenticated user elevated privileges on the SQL database, which would allow an authenticated user to modify drug libraries, add and delete users, and change user permissions. According to Smiths-Medical, physical access to the pump is required to install drug library updates.
The product does not perform any authentication for functionality that requires a provable user identity or consumes a significant amount of resources.
Link | Tags |
---|---|
https://ics-cert.us-cert.gov/advisories/ICSMA-16-306-01 | third party advisory us government resource |
http://www.securityfocus.com/bid/94630 | vdb entry third party advisory |