CVE-2017-11580

Public Exploit

Description

Blipcare Wifi blood pressure monitor BP700 10.1 devices allow memory corruption that results in Denial of Service. When connected to the "Blip" open wireless connection provided by the device, if a large string is sent as a part of the HTTP request in any part of the HTTP headers, the device could become completely unresponsive. Presumably this happens as the memory footprint provided to this device is very small. According to the specs from Rezolt, the Wi-Fi module only has 256k of memory. As a result, an incorrect string copy operation using either memcpy, strcpy, or any of their other variants could result in filling up the memory space allocated to the function executing and this would result in memory corruption. To test the theory, one can modify the demo application provided by the Cypress WICED SDK and introduce an incorrect "memcpy" operation and use the compiled application on the evaluation board provided by Cypress semiconductors with exactly the same Wi-Fi SOC. The results were identical where the device would completely stop responding to any of the ping or web requests.

Category

6.5
CVSS
Severity: Medium
CVSS 3.0 •
CVSS 2.0 •
EPSS 0.50%
Third-Party Advisory packetstormsecurity.com Third-Party Advisory github.com Third-Party Advisory seclists.org
Affected: n/a n/a
Published at:
Updated at:

References

Frequently Asked Questions

What is the severity of CVE-2017-11580?
CVE-2017-11580 has been scored as a medium severity vulnerability.
How to fix CVE-2017-11580?
To fix CVE-2017-11580, make sure you are using an up-to-date version of the affected component(s) by checking the vendor release notes. As for now, there are no other specific guidelines available.
Is CVE-2017-11580 being actively exploited in the wild?
It is possible that CVE-2017-11580 is being exploited or will be exploited in a near future based on public information. According to its EPSS score, there is a ~0% probability that this vulnerability will be exploited by malicious actors in the next 30 days.
This platform uses data from the NIST NVD, MITRE CVE, MITRE CWE, First.org and CISA KEV but is not endorsed or certified by these entities. CVE is a registred trademark of the MITRE Corporation and the authoritative source of CVE content is MITRE's CVE web site. CWE is a registred trademark of the MITRE Corporation and the authoritative source of CWE content is MITRE's CWE web site.