Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity.
The product receives input or data, but it does not validate or incorrectly validates that the input has the properties that are required to process the data safely and correctly.
Link | Tags |
---|---|
https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01 | third party advisory us government resource |
http://www.securityfocus.com/bid/106683 | third party advisory vdb entry |