CVE-2018-7518

Description

In TotalAlert Web Application in BeaconMedaes Scroll Medical Air Systems prior to v4107600010.23, an attacker with network access to the integrated web server could retrieve default or user defined credentials stored and transmitted in an insecure manner.

Category

9.8
CVSS
Severity: Critical
CVSS 3.0 •
CVSS 2.0 •
EPSS 0.25%
Third-Party Advisory us-cert.gov
Affected: ICS-CERT BeaconMedæs TotalAlert Scroll Medical Air Systems web application
Published at:
Updated at:

References

Link Tags
https://ics-cert.us-cert.gov/advisories/ICSMA-18-144-01 mitigation third party advisory us government resource

Frequently Asked Questions

What is the severity of CVE-2018-7518?
CVE-2018-7518 has been scored as a critical severity vulnerability.
How to fix CVE-2018-7518?
To fix CVE-2018-7518, make sure you are using an up-to-date version of the affected component(s) by checking the vendor release notes. As for now, there are no other specific guidelines available.
Is CVE-2018-7518 being actively exploited in the wild?
As for now, there are no information to confirm that CVE-2018-7518 is being actively exploited. According to its EPSS score, there is a ~0% probability that this vulnerability will be exploited by malicious actors in the next 30 days.
What software or system is affected by CVE-2018-7518?
CVE-2018-7518 affects ICS-CERT BeaconMedæs TotalAlert Scroll Medical Air Systems web application.
This platform uses data from the NIST NVD, MITRE CVE, MITRE CWE, First.org and CISA KEV but is not endorsed or certified by these entities. CVE is a registred trademark of the MITRE Corporation and the authoritative source of CVE content is MITRE's CVE web site. CWE is a registred trademark of the MITRE Corporation and the authoritative source of CWE content is MITRE's CWE web site.