Fujifilm FCR Capsula X/ Carbon X/ FCR XC-2, model versions CR-IR 357 FCR Carbon X, CR-IR 357 FCR XC-2, FCR-IR 357 FCR Capsula X are susceptible to a denial-of-service condition as a result of an overflow of TCP packets, which requires the device to be manually rebooted.
The product does not properly control the allocation and maintenance of a limited resource.
Link | Tags |
---|---|
https://ics-cert.us-cert.gov/advisories/ICSMA-19-113-01 | third party advisory us government resource |