The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.
Solution:
Workaround:
The product transmits sensitive or security-critical data in cleartext in a communication channel that can be sniffed by unauthorized actors.
Link | Tags |
---|---|
https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01 | third party advisory us government resource |
http://www.securityfocus.com/bid/107544 | broken link vdb entry |