CVE-2019-6540

Medtronic Conexus Radio Frequency Telemetry Protocol Cleartext Transmission of Sensitive Information

Description

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.

Remediation

Solution:

  • Medtronic has developed mitigating patches for a subset of the affected implanted cardiac device models. These patches are installed during regular office visits. Medtronic has stated that patches for additional impacted models are being developed by Medtronic and will be deployed through future updates. Patches are currently available for the following affected models: * Protecta CRT-D and implantable cardioverter defibrillators (ICDs), all models * Amplia MRI CRT-D, all models (patch available in U.S. only) * Claria MRI CRT-D, all models (patch available in U.S. only) * Compia MRI CRT-D, all models (patch available in U.S. only) * Visia AF MRI ICD, all models * Visia AF ICD, all models * Brava CRT-D, all models                                             * Evera MRI ICD, all models * Evera ICD, all models * Mirro MRI ICD, all models * Primo MRI ICD, all models * Viva CRT-D, all models

Workaround:

  • Medtronic has applied additional controls for monitoring and responding to improper use of the Conexus telemetry protocol by the affected implanted cardiac devices. Additional mitigations are being developed and will be deployed through future updates, assuming regulatory approval. Medtronic recommends that users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should: * Maintain good physical control over home monitors and programmers. * Use only home monitors, programmers, and implantable devices obtained directly from your healthcare provider or a Medtronic representative to ensure integrity of the system. * Do not connect unapproved devices to home monitors and programmers through USB ports or other physical connections. * Only use programmers to connect and interact with implanted devices in physically controlled hospital and clinical environments. * Only use home monitors in private environments such as a home, apartment, or otherwise physically controlled environment. * Report any concerning behavior regarding these products to your healthcare provider or a Medtronic representative. Medtronic has released additional patient focused information https://www.medtronic.com/security .

Category

6.5
CVSS
Severity: Medium
CVSS 3.1 •
CVSS 2.0 •
EPSS 0.02%
Third-Party Advisory us-cert.gov
Affected: Medtronic Conexus Radio Frequency Telemetry Protocol
Affected: Medtronic MyCareLink Monitor
Affected: Medtronic CareLink Monitor
Affected: Medtronic CareLink 2090 Programmer
Affected: Medtronic Amplia CRT-D
Affected: Medtronic Claria CRT-D
Affected: Medtronic Compia CRT-D
Affected: Medtronic Concerto CRT-D
Affected: Medtronic Concerto II CRT-D
Affected: Medtronic Consulta CRT-D
Affected: Medtronic Evera ICD
Affected: Medtronic Maximo II CRT-D
Affected: Medtronic Maximo II ICD
Affected: Medtronic Mirro ICD
Affected: Medtronic Nayamed ND ICD
Affected: Medtronic Primo ICD
Affected: Medtronic Protecta ICD, Protecta CRT-D
Affected: Medtronic Secura ICD
Affected: Medtronic Virtuoso ICD
Affected: Medtronic Virtuoso II ICD
Affected: Medtronic Visia AF ICD
Affected: Medtronic Viva CRT-D
Affected: Medtronic Brava CRT-D
Affected: Medtronic Mirro MRI ICD
Published at:
Updated at:

References

Link Tags
https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01 third party advisory us government resource
http://www.securityfocus.com/bid/107544 broken link vdb entry

Frequently Asked Questions

What is the severity of CVE-2019-6540?
CVE-2019-6540 has been scored as a medium severity vulnerability.
How to fix CVE-2019-6540?
To fix CVE-2019-6540: Medtronic has developed mitigating patches for a subset of the affected implanted cardiac device models. These patches are installed during regular office visits. Medtronic has stated that patches for additional impacted models are being developed by Medtronic and will be deployed through future updates. Patches are currently available for the following affected models: * Protecta CRT-D and implantable cardioverter defibrillators (ICDs), all models * Amplia MRI CRT-D, all models (patch available in U.S. only) * Claria MRI CRT-D, all models (patch available in U.S. only) * Compia MRI CRT-D, all models (patch available in U.S. only) * Visia AF MRI ICD, all models * Visia AF ICD, all models * Brava CRT-D, all models                                             * Evera MRI ICD, all models * Evera ICD, all models * Mirro MRI ICD, all models * Primo MRI ICD, all models * Viva CRT-D, all models
Is CVE-2019-6540 being actively exploited in the wild?
As for now, there are no information to confirm that CVE-2019-6540 is being actively exploited. According to its EPSS score, there is a ~0% probability that this vulnerability will be exploited by malicious actors in the next 30 days.
What software or system is affected by CVE-2019-6540?
CVE-2019-6540 affects Medtronic Conexus Radio Frequency Telemetry Protocol, Medtronic MyCareLink Monitor, Medtronic CareLink Monitor, Medtronic CareLink 2090 Programmer, Medtronic Amplia CRT-D, Medtronic Claria CRT-D, Medtronic Compia CRT-D, Medtronic Concerto CRT-D, Medtronic Concerto II CRT-D, Medtronic Consulta CRT-D, Medtronic Evera ICD, Medtronic Maximo II CRT-D, Medtronic Maximo II ICD, Medtronic Mirro ICD, Medtronic Nayamed ND ICD, Medtronic Primo ICD, Medtronic Protecta ICD, Protecta CRT-D, Medtronic Secura ICD, Medtronic Virtuoso ICD, Medtronic Virtuoso II ICD, Medtronic Visia AF ICD, Medtronic Viva CRT-D, Medtronic Brava CRT-D, Medtronic Mirro MRI ICD.
This platform uses data from the NIST NVD, MITRE CVE, MITRE CWE, First.org and CISA KEV but is not endorsed or certified by these entities. CVE is a registred trademark of the MITRE Corporation and the authoritative source of CVE content is MITRE's CVE web site. CWE is a registred trademark of the MITRE Corporation and the authoritative source of CWE content is MITRE's CWE web site.