Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
The product contains a hard-coded password, which it uses for its own inbound authentication or for outbound communication to external components.
The product transmits sensitive or security-critical data in cleartext in a communication channel that can be sniffed by unauthorized actors.
Link | Tags |
---|---|
https://www.us-cert.gov/ics/advisories/icsma-20-170-01 | not applicable |
https://us-cert.cisa.gov/ics/advisories/icsma-20-170-02 | third party advisory |