CVE-2022-3027

Contec Health CMS8000

Description

The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attempts connecting to the malicious SSID, the device can be exploited to write arbitrary files or display incorrect information.

Remediation

Workaround:

  • Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information. The following mitigations could assist in reducing the risk for exploitation of vulnerabilities: Disabling UART functionality at the CPU level Enforcing unique device authentication before granting access to the terminal / bootloader Where possible, enforcing secure boot. Tamper stickers on the device casing to indicate when a device has been opened

Category

5.7
CVSS
Severity: Medium
CVSS 3.1 •
EPSS 0.03%
Third-Party Advisory cisa.gov
Affected: Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor
Published at:
Updated at:

References

Link Tags
https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01 us government resource third party advisory mitigation

Frequently Asked Questions

What is the severity of CVE-2022-3027?
CVE-2022-3027 has been scored as a medium severity vulnerability.
How to fix CVE-2022-3027?
As a workaround for remediating CVE-2022-3027: Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information. The following mitigations could assist in reducing the risk for exploitation of vulnerabilities: Disabling UART functionality at the CPU level Enforcing unique device authentication before granting access to the terminal / bootloader Where possible, enforcing secure boot. Tamper stickers on the device casing to indicate when a device has been opened
Is CVE-2022-3027 being actively exploited in the wild?
As for now, there are no information to confirm that CVE-2022-3027 is being actively exploited. According to its EPSS score, there is a ~0% probability that this vulnerability will be exploited by malicious actors in the next 30 days.
What software or system is affected by CVE-2022-3027?
CVE-2022-3027 affects Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor.
This platform uses data from the NIST NVD, MITRE CVE, MITRE CWE, First.org and CISA KEV but is not endorsed or certified by these entities. CVE is a registred trademark of the MITRE Corporation and the authoritative source of CVE content is MITRE's CVE web site. CWE is a registred trademark of the MITRE Corporation and the authoritative source of CWE content is MITRE's CWE web site.
© 2025 Under My Watch. All Rights Reserved.