CVE-2022-38069

Contec Health CMS8000

Description

Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters

Remediation

Workaround:

  • Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information. The following mitigations could assist in reducing the risk for exploitation of vulnerabilities: Disabling UART functionality at the CPU level Enforcing unique device authentication before granting access to the terminal / bootloader Where possible, enforcing secure boot. Tamper stickers on the device casing to indicate when a device has been opened

Category

4.3
CVSS
Severity: Medium
CVSS 3.1 •
EPSS 0.06%
Third-Party Advisory cisa.gov
Affected: Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor
Published at:
Updated at:

References

Link Tags
https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01 mitigation third party advisory us government resource

Frequently Asked Questions

What is the severity of CVE-2022-38069?
CVE-2022-38069 has been scored as a medium severity vulnerability.
How to fix CVE-2022-38069?
As a workaround for remediating CVE-2022-38069: Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information. The following mitigations could assist in reducing the risk for exploitation of vulnerabilities: Disabling UART functionality at the CPU level Enforcing unique device authentication before granting access to the terminal / bootloader Where possible, enforcing secure boot. Tamper stickers on the device casing to indicate when a device has been opened
Is CVE-2022-38069 being actively exploited in the wild?
As for now, there are no information to confirm that CVE-2022-38069 is being actively exploited. According to its EPSS score, there is a ~0% probability that this vulnerability will be exploited by malicious actors in the next 30 days.
What software or system is affected by CVE-2022-38069?
CVE-2022-38069 affects Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor.
This platform uses data from the NIST NVD, MITRE CVE, MITRE CWE, First.org and CISA KEV but is not endorsed or certified by these entities. CVE is a registred trademark of the MITRE Corporation and the authoritative source of CVE content is MITRE's CVE web site. CWE is a registred trademark of the MITRE Corporation and the authoritative source of CWE content is MITRE's CWE web site.
© 2025 Under My Watch. All Rights Reserved.