No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator

Description

There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.

Category

9.3
CVSS
Severity: Critical
CVSS 3.1 •
EPSS 0.03%
Affected: Baxter Life2000 Ventilation System
Published at:
Updated at:

References

Frequently Asked Questions

What is the severity of CVE-2024-9832?
CVE-2024-9832 has been scored as a critical severity vulnerability.
How to fix CVE-2024-9832?
To fix CVE-2024-9832, make sure you are using an up-to-date version of the affected component(s) by checking the vendor release notes. As for now, there are no other specific guidelines available.
Is CVE-2024-9832 being actively exploited in the wild?
As for now, there are no information to confirm that CVE-2024-9832 is being actively exploited. According to its EPSS score, there is a ~0% probability that this vulnerability will be exploited by malicious actors in the next 30 days.
What software or system is affected by CVE-2024-9832?
CVE-2024-9832 affects Baxter Life2000 Ventilation System.
This platform uses data from the NIST NVD, MITRE CVE, MITRE CWE, First.org and CISA KEV but is not endorsed or certified by these entities. CVE is a registred trademark of the MITRE Corporation and the authoritative source of CVE content is MITRE's CVE web site. CWE is a registred trademark of the MITRE Corporation and the authoritative source of CWE content is MITRE's CWE web site.